Pyrometer Readings Low

From the choices you have made the problem you are experiencing
is most likely to be caused by problems with :

The most likely causes are: Remedial Actions after overhaul are:
  • Purge Air set too high
  • Lens dirty

  • Fibre optic cable damaged

  • Instrument DC power supply low
  • Incorrect range unit fitted
  • Fibre optic cable replaced in house
  • Reduce Purge air
  • Clean lens with soft dry lint free cloth
  • Requires repair contact BH-F
  • Adjust / Replace

  • Re fit correct range unit
  • Requires factory calibration refer to BH-F